There is a problem with one APD in one of the submodules currently under test: Need to figure out if the failure is real or just a cabling problem or something like that.
We decided to have another exchange of submodules by Christoph in Bochum in the near future. Christoph will also bring HV distributor boards
Drifting of set voltages on the HV distributor boards is still not fully understood. Using a 100 (or 22) MOhm instead of the 10 MOhm one currently used seems to lower the drift and is therefore suggested to use by Christoph
A casting compound (like Elastosil we use for read out unit casting) on the resistors may help in preventing surface currents
One one submodule there seems to be a problem with noise introduction from a temperature sensor to the respective channel (need to take a closer look at it, never observed any problems like that in Bochum)
SADC Crates:
Fritz-Herbert asked what power (voltage, current) needs to be supplied to the SADC crate power supplies (+ 12 V only?, what current?)