Minutes of the forward endcap EMC eZuce SRN meeting, 31th of July 2017, 14:00-14:30

Presence:

  • Basel: Werner
  • Bochum: Cathrina, Claudius, Fritz-Herbert, Malte, Matthias, Miriam
  • Bonn: excused
  • Groningen: -
  • Stockholm: -
  • Uppsala: -
  • Giessen: -
  • GSI: Holger

Preamplifier

  • Malte presented gain measurements of 15 preamplifiers for each of the three different preamplifier groups: Red label (except 2 preamps) and Blue label are ok and consistent. White label show large variations with mostly larger gain than expected. The deviations were confirmed by re-measurements with 270pF charge injection instead of 100pF.
  • Werner suggested to measure with a standard setup of 1pF charge injection. This was not done, because errors due to capacities of the cables are to be avoided.
  • Malte will send a list of the serial numbers to Werner and Nathalie. They will cross check with the measurements done at Basel.
  • Werner will ask Nathalie to provide the data from the validation measurements of the preamps at Basel.

-- FritzHerbertHeinsius - 31 Jul 2017