The THMPis has been developed to monitor the temperature and humidity inside the electromagnetic calorimeter (EMC) of the PANDA target spectrometer. Therefore it has 64 channels and is designed to read-out read out PT100 temperature sensors and HIH-4000humidity sensors. This device is has been constructed to operate in a temperature range from -30 °C to +30 °C. To cover the whole temperature range of the EMC the measurement range will be also be from -30 °C to +30 °C with a resolution of 0.05 K. The relative humidity (RH) can be measured in a range from 0 to 100% RH with an accuracy of 3.5%. For the read-out of this device a CAN interface is foreseen.
The THMP should fulfill the environmental requirements of operating in the cooled area of the EMC. Different tests were carried out to verify this.
-- FlorianFeldbauer- , 17 Nov 2009 PatrickFriedel - 30 Nov 2009
I | Attachment![]() | Action | Size | Date | Who | Comment |
---|---|---|---|---|---|---|
![]() | ThmpReport_Rev1.0.pdf | manage | 1980.1 K | 17 Nov 2009 - 14:03 | FlorianFeldbauer | Documentation of the development of the THMP |